欧美一区二区三区婷婷五,亚洲qing色中文字幕,久久久久久精品国产夫妻,玩弄人妻少妇500系列网址

本站熱搜:氧化鋯氧量分析儀,盤裝式氧量分析儀,電磁流量計,WSS雙金屬溫度計
  • 技術文章ARTICLE

    您當前的位置:首頁 > 技術文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發(fā)布時間: 2024-12-13  點擊次數(shù): 1368次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


產(chǎn)品中心 Products
在線客服 聯(lián)系方式 二維碼

服務熱線

0550-2409223

掃一掃,關注我們

皖公網(wǎng)安備 34118102000850號

久久久精品免费免费直播| 老张和老李互相换女h| 国产美女 91 在线播放| 国产精品久久久久久小说| 日本不卡三区| 国产免国产免费| 新天堂网www在线视频| 精品人妻大屁股白浆无码| 久久精品加勒比中文字幕| 亚洲欧美激情精品一区二区 | 亚洲天堂av一区二区三区| 暴露放荡的娇妻| 国产精品一区二区蜜臀av| 国产色婷婷视频在线观看| 久久综合AV色老头免费观看| 久久久久久久久久久性视频| 成人精品视频一区二区三区不卡| 免费脚交足在线播放视频| 777婷婷天堂综合区色吧| 午夜剧场韩国| 欧洲无码精品A码无人区| 成全视频在线观看在线播放高清 | gogogo免费高清人体摄影 | asian北京少妇pics| 暴躁妹妹csgo暴躁老外| 午夜性无码专区| 天天夜i日日清莫一97| 99精品人妻少妇一区二区| 日韩高清电影一区二区三区| 国产成人精品999在线观看| 久久变态刺激另类sm按摩| 中文字幕av午夜福利片| 亚洲色偷偷综合亚洲AV| 亚洲最色最色最色的视频| www国产在线| 757午夜视频国产精品| 国产欧美日韩精品丝袜高跟鞋| 老司机深夜福利在线观看| av无码久久久久久不卡网站| 99国产精品人妻无码| 日韩亚洲欧美一区二区三区|